;************************** BLIS INTERFACE CLIENT CONFIGURATION FILE ********************************************** | |
;This is the main configuration file for C4G BLIS Interface Client. | |
;BLIS Interface Client is a very lightweight client for Interfacing BLIS with external systems using Protocols like | |
;RS232 Serial Port,TCP/IP, HTTP | |
; | |
; | |
; C4G BLIS Equipment Interface Client | |
; | |
; Project funded by PEPFAR | |
; | |
; Philip Boakye - Team Lead | |
; Patricia Enninful - Technical Officer | |
; Stephen Adjei-Kyei - Software Developer | |
; | |
; | |
;Author: Stephen Adjei-Kyei <[email protected]> | |
;****************************************************************************************************************** | |
;****************************************************** | |
;The source from which information will be received | |
;Supported protocols are (RS232,TCP/IP, MSACCESS,HTTP,TEXT) | |
;******************************************************* | |
[FEED SOURCE] | |
TEXT | |
;*************************************************** | |
; | |
;Configuration settings for Microsoft Access database connection | |
; | |
;*************************************************** | |
[MSACCESS CONFIGURATIONS] | |
;The ODBC Datasouce name | |
DATASOURCE = PentraCon | |
;How many days old test data should be imported into msacess database. | |
;Default is 1 day old test data | |
DAYS = 3 | |
;******************************************************** | |
; | |
;configuration settings for serial port connection | |
; | |
;******************************************************** | |
[RS232 CONFIGURATIONS] | |
;The comm port on which to connect | |
COMPORT = COM21 | |
;The speed of connection.eg (4800,9600,38400,57600,115200) | |
BAUD_RATE = 9600 | |
;How should data be checked for errors. Available options are (Odd,Even,None) | |
PARITY = None | |
;Choose the time between character transmission for your serial device. Available options are (1,2) | |
STOP_BITS = 1 | |
;Choose the number of data bits used by your serial device to represent characters. Available options are (7,8) | |
DATA_BITS = 8 | |
;Should the data transfer be appended with the Newline(/n) character (Yes/No) | |
APPEND_NEWLINE = No | |
;Should the data transfer be appended with the Enter(/r) character (Yes/No) | |
APPEND_CARRIAGE_RETURN = No | |
;*************************************************** | |
; | |
;Configuration settings for TCP/IP connection | |
; | |
;*************************************************** | |
[TCP/IP CONFIGURATIONS] | |
;Port on which to listen for data | |
PORT = 5150 | |
;equipment IP address if applicable | |
EQUIPMENT_IP = 192.168.6.24 | |
;How should the application connect to analyser. options (server/client) | |
MODE = server | |
;This setting is used only if MODE=client | |
;Reset client Connection when server sends "End of Transmission (EOT)". Options (yes/no) | |
CLIENT_RECONNECT = no | |
;*************************************************** | |
; | |
;configuration for flat file connections | |
; | |
;*************************************************** | |
[TEXT] | |
;The parent directory where all files will be located | |
BASE_DIRECTORY = /home/reaper/HI/ | |
;Directory to save worklist | |
INPUT_DIRECTORY = Input Worklist | |
;Directory where worklist is output | |
OUTPUT_DIRECTORY = Output Worklist | |
;Should sub directories be opened for files? (OPtions are yes/no) | |
USE_SUB_DIRECTORIES = no | |
;************************************************** | |
;DATE FORMAT TABLE | |
;************************************************** | |
; Symbol Meaning Presentation Example | |
; G era designator Text AD | |
; y year Number 2009 | |
; M month in year Text & Number July & 07 | |
; d day in month Number 10 | |
; h hour in am/pm (1-12) Number 12 | |
; H hour in day (0-23) Number 0 | |
; m minute in hour Number 30 | |
; s second in minute Number 55 | |
; S millisecond Number 978 | |
; E day in week Text Tuesday | |
; D day in year Number 189 | |
; F day of week in month Number 2 (2nd Wed in July) | |
; w week in year Number 27 | |
; W week in month Number 2 | |
; a am/pm marker Text PM | |
; k hour in day (1-24) Number 24 | |
; K hour in am/pm (0-11) Number 0 | |
; z time zone Text Pacific Standard Time | |
; ' escape for text Delimiter (none) | |
; ' single quote Literal ' | |
;sub directories naming format. Example ddMMyy. Only used when USE_SUB_DIRECTORIES = yes | |
SUB_DIRECTORY_FORMAT = ddMMyy | |
;File name | |
FILE_NAME = myworklist | |
;Files naming format | |
FILE_NAME_FORMAT = ddMMyy | |
;File extension used. Example txt,csv,exp,evl | |
FILE_EXTENSION = astm | |
;File value separator used. Options (TAB,COMMA,COLON,SEMI-COLON,SPACE,VERTICAL-LINE) | |
FILE_SEPERATOR = VERTICAL-LINE | |
;*************************************************** | |
; | |
;configuration of BLIS server | |
; | |
;*************************************************** | |
[BLIS CONFIGURATIONS] | |
;BLIS URL local | |
;BLIS_URL = http://192.168.56.1:4001/ | |
;username | |
;BLIS_USERNAME = equipment | |
;BLIS_PASSWORD = equipment | |
;Live server | |
BLIS_URL = http://192.168.1.45:4001/ | |
BLIS_USERNAME = Selectra Junior | |
BLIS_PASSWORD = Selectra Junior | |
;********************************************************************************************************************* | |
; | |
;Equipment Name section | |
;This section contains the list of all supported equipment this version of BLIS Interface Client can connect. | |
;Please note that if your equipment is not listed here you cannot connect with this Version of BLIS Interface Client. | |
;To use any of the equipment listed here just uncomment that line with the equipment name. | |
;Only 1 equipment can be connected at a time. To connect to more equipment at a time from the same PC, create multiple instances of BLIS Interface Client and set their configurations accordingly | |
; | |
;PLEASE TAKE NOTE OF THE SUPPORTED FEED SOURCE ASSIGNED TO EACH EQUIPMENT AND SET THEM ACCORDINGLY AT THE FEED SOURCE SECTION | |
; | |
;********************************************************************************************************************** | |
[EQUIPMENT] | |
;ABX Pentra 60C+ | |
;Supported Feed Source = MSACCESS | |
;2-way communication | |
;ABX MICROS 60 | |
;Supported Feed Source = RS232 | |
;1-way communication | |
;ABX Pentra 80 | |
;Supported Feed Source = RS232 | |
;2-way communication | |
;Selectra Junior | |
;Supported Feed Source = RS232 | |
;1-way communication | |
;Flexor E | |
;Supported Feed Source = RS232 | |
;1-way communication | |
;MINDRAY BC 3600 | |
;Supported Feed Source = RS232 | |
;1-way communication | |
;BT3000 Plus-Chameleon | |
;Supported Feed Source = TCP/IP and RS232 | |
;1-way communication | |
;Mindray BS-200E | |
;Supported Feed Source = TCP/IP | |
;2-way communication | |
;BT3000 Plus | |
;Supported Feed Source = TCP/IP | |
;1-way communication | |
;SYSMEX XS-500i | |
;Supported Feed Source = TCP/IP | |
;1-way communication | |
;Flexor Junior | |
;Supported Feed Source = TCP/IP | |
;1-way communication | |
;GeneXpert | |
;Supported Feed Source = TCP/IP | |
;2-way communication | |
;SYSMEX XT-2000i | |
;Supported Feed Source = TCP/IP | |
;1-way communication | |
;BD FACSCalibur | |
;Supported Feed Source = TEXT | |
;1-way communication | |
Humastar 100 | |
;Supported Feed Source = TEXT | |
;1-way communication | |
;SUPPORTED BUT NOT TESTED | |
;Cobas AmpliPrep | |
;Supported Feed Source = TCP/IP | |
;2-way communication | |
;*********************************************************************************************************************** | |
;Other miscellaneous application configurations | |
;*********************************************************************************************************************** | |
[MISCELLANEOUS] | |
; options are yes/no | |
ENABLE_LOG = yes | |
;Option to enable writing messages received from analyzer to file. Options are yes/no | |
WRITE_TO_FILE = yes | |
;How many days old data to pull from BLIS. 0: is for current day, 1: for yesterday, 2: for 2 days ago, etc | |
;This configuration is only applicable to ABX Pentra 60C+ | |
POOL_DAY = 0 | |
;Time interval in Seconds to wait and fetch for data. Default is 30 seconds | |
POOL_INTERVAL = 30 | |
; Set whether automatic data retrieval is enabled or not. Options are yes/no | |
ENABLE_AUTO_POOL = yes | |
;Set whether LAB is using BLIS auto generated specimen IDs. Options are yes/no | |
AUTO_SPECIMEN_ID = yes |